Evaluation of The Clinical Performance of LMA Proseal versus Blockbuster LMA in Adult Patients Undergoing General Anesthesia: a Randomized Interventional Comparative Study

Authors

  • Basavaraj Ryavanaki Vardhaman Mahavir Medical College and Safdarjung Hospital, New Delhi, India
  • Arushi Gupta Vardhaman Mahavir Medical College and Safdarjung Hospital, New Delhi, India
  • Ameeta Sahni Vardhaman Mahavir Medical College and Safdarjung Hospital, New Delhi, India
  • Khushboo Mehta Vardhaman Mahavir Medical College and Safdarjung Hospital, New Delhi, India

DOI:

https://doi.org/10.21776/

Keywords:

Blockbuster LMA, General anaesthesia, Oropharyngeal leak pressure, LMA ProSeal, Supraglottic airway device

Abstract

Background: The oropharyngeal Leak pressure (OLP) and the minimum effective cuff volume are the two most important determinants of the clinical performance of any supraglottic airway (SGA) device. This study aims to evaluate the clinical performance of the Blockbuster LMA compared with the LMA ProSeal.

Methods: One-Hundred patients (18-65 years) posted for elective surgeries under general anesthesia were prospectively randomized into group P (LMA ProSeal) and Group B (Blockbuster LMA). The primary outcome was to measure the OLP at 5 and 30 min after device insertion, and, secondarily, to compare the minimum effective cuff volume to prevent leak, ease and time of insertion, number of attempts, leak fraction, anatomical position of the devices, and adverse respiratory events, if any.

Result: The OLP of LMA ProSeal was lower than Blockbuster LMA (28.14 ± 2.79 cmH2O versus 32.62 ± 2.66 cmH2O at 5 minutes and 29.28 ± 3.1 cmH2O versus 33.82 ± 2.55 cmH2O at 30 minutes). The cuff pressure exerted by the Blockbuster LMA at the minimum effective volume to prevent leak was higher than that of the LMA ProSeal. The first attempt success rate and Brimacombe score were higher in Blockbuster LMA. The airway-sealing quality and the incidence of postoperative adverse respiratory events were similar between the two groups.

Conclusion: Blockbuster LMA has a higher OLP, a better first-attempt success rate, and an anatomical position than LMA ProSeal, with higher pressure at the minimum effective volume of cuff inflation to prevent leaks, along with the added advantage of acting as a conduit for intubation.

References

Norris AM, Hardman JG, Asai T. A firm foundation for progress in airway management. Br J Anaesth. 2011;106(5):613-616. doi:10.1093/bja/aer088

Brain AI. The laryngeal mask--a new concept in airway management. Br J Anaesth. 1983;55(8):801-805. doi:10.1093/bja/55.8.801

LMA® ProSealTM Airway. Teleflex® | LMA®. Published December 19, 2013. Accessed November 1, 2023. https://www.lmaco.com/products/lma%C2%AE-proseal%E2%84%A2-airway

Miller DM. A proposed classification and scoring system for supraglottic sealing airways: a brief review. Anesth Analg.

;99(5):1553-1559. doi:10.1213/01.ANE.0000134798.00069.2B

Brain AIJ, Verghese C, Strube PJ. The LMA ‘ProSeal’—a laryngeal mask with an oesophageal vent. Br J Anaesth. 2000;84(5):650-654. doi:10.1093/bja/84.5.650

BlockBusterTM Laryngeal Mask Airway. Accessed June 15, 2023. https://tuoren.in/product/blockbuster-TM-laryngeal-mask-airway

Joshi R, Rudingwa P, Kundra P, Panneerselvam S, Mishra SK. Comparision of Ambu AuraGainTM and LMA® ProSeal in children under controlled ventilation. Indian J Anaesth. 2018;62(6):455-460. doi:10.4103/ija.IJA_86_18

Endigeri A, Ganeshnavar A, Varaprasad B, Shivanand Y, Ayyangouda B. Comparison of success rate of BlockBuster® versus Fastrach® LMA as conduit for blind endotracheal intubation: A prospective randomised trial. Indian J Anaesth. 2019;63(12):988-994. doi:10.4103/ija.IJA_396_19

Guleria A, Devanshi, Sood P, Badhan A, Singha D, Singh J. European Journal of Molecular & Clinical Medicine THE BLOCKBUSTER LMA: A Prospective Cohort Study to evaluate the safety and efficacy of the device. Published online May 29, 2023.

Asai T, Howell TK, Koga K, Morris S. Appropriate size and inflation of the laryngeal mask airway. Br J Anaesth. 1998;80(4):470-474. doi:10.1093/bja/80.4.470

Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993;76(2):457.

Park HP. Supraglottic airway devices: more good than bad. Korean J Anesthesiol. 2019;72(6):525-526. doi:10.4097/kja.19417

Maitra S, Khanna P, Baidya DK. Comparison of laryngeal mask airway Supreme and laryngeal mask airway Pro-Seal for controlled ventilation during general anaesthesia in adult patients: Systematic review with meta-analysis. Eur J Anaesthesiol EJA. 2014;31(5):266. doi:10.1097/01.EJA.0000435015.89651.3d

Lopez-Gil M, Brimacombe J, Keller C. A comparison of four methods for assessing oropharyngeal leak pressure with the laryngeal mask airway (LMA) in paediatric patients. Paediatr Anaesth. 2001;11(3):319-321. doi:10.1046/j.1460-9592.2001.00649.x

Singh K, Gurha P. Comparative evaluation of Ambu AuraGainTM with ProSealTM laryngeal mask airway in patients undergoing laparoscopic cholecystectomy. Indian J Anaesth. 2017;61(6):469-474. doi:10.4103/ija.IJA_163_17

Gasteiger L, Brimacombe J, Perkhofer D, Kaufmann M, Keller C. ORIGINAL ARTICLE: Comparison of guided insertion of the LMA ProSealTM vs the i-gelTM. Anaesthesia. 2010;65(9):913-916. doi:10.1111/j.1365-2044.2010.06422.x

Brimacombe J, Keller C, Boehler M, Pühringer F. Positive Pressure Ventilation with the ProSeal Versus Classic Laryngeal Mask Airway: A Randomized, Crossover Study of Healthy Female Patients. Anesth Analg. 2001;93(5):1351. doi:10.1097/00000539-200111000-00064

Selvin CC, Singariya G, Bihani P, Kamal M, Paliwal N, Ujwal S. Comparison of oropharyngeal leak pressure of I-gelTM and BlockbusterTM laryngeal mask airway in anaesthetized pediatric patients. Anesth Pain Med. 2023;18(1):51-56. doi:10.17085/apm.22209

Brimacombe J, Keller C, Pühringer F. Pharyngeal mucosal pressure and perfusion: a fiberoptic evaluation of the posterior pharynx in anesthetized adult patients with a modified cuffed oropharyngeal airway. Anesthesiology. 1999;91(6):1661-1665. doi:10.1097/00000542-199912000-00018

Marjot R. Pressure exerted by the laryngeal mask airway cuff upon the pharyngeal mucosa. Br J Anaesth. 1993;70(1):25-29. doi:10.1093/bja/70.1.25

Seet E, Rajeev S, Firoz T, et al. Safety and efficacy of laryngeal mask airway Supreme versus laryngeal mask airway ProSeal: a randomized controlled trial. Eur J Anaesthesiol. 2010;27(7):602-607. doi:10.1097/eja.0b013e32833679e3

Jadhav PA, Dalvi NP, Tendolkar BA. I-gel versus laryngeal mask airway-Proseal: Comparison of two supraglottic airway devices in short surgical procedures. J Anaesthesiol Clin

Pharmacol. 2015;31(2):221-225. doi:10.4103/0970-9185.155153

Chauhan G, Nayar P, Seth A, Gupta K, Panwar M, Agrawal N. Comparison of clinical performance of the I-gel with LMA proseal. J Anaesthesiol Clin Pharmacol. 2013;29(1):56-60. doi:10.4103/0970-9185.105798

Sood S, Chahar S, Thakur A, Gupta M, Saxena A, Subramanian S. Comparison and evaluation of single-use LMA supreme versus the reusable proseal LMA in paralyzed patients undergoing surgery with controlled ventilation. J Anaesthesiol Clin Pharmacol. 2020;36(4):494. doi:10.4103/joacp.JOACP_298_18

Tham HM, Tan SM, Woon KL, Zhao YD. A comparison of the Supreme laryngeal mask airway with the Proseal laryngeal mask airway in anesthetized paralyzed adult patients: a randomized crossover study. Can J Anaesth J Can Anesth. 2010;57(7):672-678. doi:10.1007/s12630-010-9312-6

Published

2026-06-12

Issue

Section

Research Articles

How to Cite

Evaluation of The Clinical Performance of LMA Proseal versus Blockbuster LMA in Adult Patients Undergoing General Anesthesia: a Randomized Interventional Comparative Study. (2026). Journal of Anaesthesia and Pain, 6(2). https://doi.org/10.21776/